Quick summary: Zeus® Infinity® Empowered. Date: 2025-01-07.
Zeus® Infinity® Empowered
2025-01-07
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Zeus® Infinity® Empowered. Date: 2025-01-07.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
In the course of service activities, it was found that CO2 readings were slightly increased. In some cases, the ISO accuracy was no longer achieved. The tolerance check is intended to ensure that the zeus ie will measure all gases within the specified accuracy until the next annual check. Draeger has updated the instructions for use of the Zeus® ie anesthesia device regarding the test tolerance of the CO2 measurement. Recall start date: December 23, 2024
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.