Zenition 50
2025-10-15
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Zenition 50. Date: 2025-10-15.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Philips has been notified by a supplier about an issue affecting integrated chips used in the image intensifier television (IITV) of the Zenition 50 systems. Due to contamination in the bonding layer of the chips on the PCBA (printed circuit board assembly) there could be corrosion of the bonding layer, potentially resulting in intermittent or permanent connection loss of the integrated chips. The issue may result in loss of imaging functionality, or clinically unusable images due to reduced image quality (blurred, reduced image size).
What you should do
- Contact the manufacturer if you require additional information.