BrandK2M, INC
ProductYukon Lock Shaft Inserter
IssueMedical devices - Performance
Recall classType II
IdentificationRA-64259

Affected products

ProductYukon Lock Shaft Inserter
LotMXGD-444564
Model7601-90186
Serial numberMXGD-444564
DescriptionYukon Lock Shaft Inserter

Issue

Stryker received reports of Yukon Lock Shaft Inserters, lot MXGD, having difficulty threading into screws and/or thread deformation observed on the inserter. Further investigation identified the thread deformation to be the result of a manufacturing non-conformance. No adverse events have been reported for this issue. Recall start date: June 3, 2022

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare
CategoryHealth products - Medical devices - Orthopaedics
CompaniesK2m, Inc 600 Hope Parkway Se, Leesburg, Virginia, United States, 20175
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-64259
Background{}
CompaniesK2m, Inc 600 Hope Parkway Se, Leesburg, Virginia, United States, 20175
AudiencesHealthcare
Source updated2022-06-13