Public recall detail

Xpert EV Assay

medium
Recall Safety notice
Quick summary

Quick summary: Xpert EV Assay.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: Xpert EV Assay.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandCepheid
ProductXpert EV Assay
IssueMedical devices - Performance
Recall classType II
IdentificationRA-64561

Affected products

ProductXpert EV Assay
Lot1000408741/22001
ModelGXEV-100N-10
Serial number1000408741/22001
DescriptionXpert EV Assay

Issue

On 27-Jul-2022 a non-conformance investigation was initiated for an issue related to a component of the test. In early August 2022, the correlation was made that this component issue may have resulted in customer complaints of invalid and false negative test results that was being investigated separately. The investigation showed that the beads used in the tests caused low sample processing control/internal control endpoints, which could lead to false negative and invalid test results. Recall start date: August 25, 2022

What you should do

  • Discontinue use of affected product, complete response form and dispose of remaining inventory.
Additional official information
AudienceHealthcare
CategoryHealth products - Medical devices
CompaniesCepheid 904 Caribbean Drive, Sunnyvale, California, United States, 94089
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-64561
Background{}
CompaniesCepheid 904 Caribbean Drive, Sunnyvale, California, United States, 94089
AudiencesHealthcare
Source updated2022-09-20

Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.