Quick summary: Xpert® CT/NG. Date: 2025-10-24.
Xpert® CT/NG
2025-10-24
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Xpert® CT/NG. Date: 2025-10-24.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Cepheid has become aware of a rare genetic mutation event between Neisseria gonorrhoeae and Neisseria meningitidis that caused some Neisseria gonorrhoeae strains to lack target sites for the Cepheid Xpert® CT/NG test, leading to false negative results. Cepheid recommends that laboratories and clinics using the Xpert® CT/NG test as their molecular method for the detection of N. gonorrhoeae monitor for decreases in the positivity rate and ensure that discordant laboratory results (e.g. testing negative on the Xpert® CT/NG test but positive for NG by culture/microscopy or testing negative on Xpert® CT/NG test but positive for NG using another NAAT) be investigated further and reported to cepheid.
What you should do
- Contact the manufacturer if you require additional information.