Quick summary: Wireless Battery Module. Date: 2024-08-01.
Wireless Battery Module
2024-08-01
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Wireless Battery Module. Date: 2024-08-01.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Baxter Corporation is issuing an urgent medical device correction for the Wireless Battery Modules (WBMs) used with Spectrum V8 and Spectrum IQ infusion pumps. Baxter is releasing a WBM software update that addresses a condition where Spectrum WBM may fail to autodocument infusion status information back to hospital's electronic medical record (EMR) system. This software issue only occurs when Spectrum V8 or Spectrum IQ infusion system is integrated with the hospital's EMR system. Recall start date: July 22, 2024
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.