Wavewriter Alpha Implantable Pulse Generator Kit

2024-08-14

medium
Recall Safety notice
Quick summary

Quick summary: Wavewriter Alpha Implantable Pulse Generator Kit. Date: 2024-08-14.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: Wavewriter Alpha Implantable Pulse Generator Kit. Date: 2024-08-14.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandBoston Scientific Neuromodulation Corporation
ProductWavewriter Alpha Implantable Pulse Generator Kit
IssueMedical devices - Performance
Recall classType II
IdentificationRA-75920

Affected products

ProductWavewriter Alpha Implantable Pulse Generator Kit
LotMore than 10 numbers, contact manufacturer.
ModelSC-1232
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionWavewriter Alpha Implantable Pulse Generator Kit
ProductWavewriter Alpha 16 Implantable Pulse Generator Kit
LotMore than 10 numbers, contact manufacturer.
ModelSC-1216
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionWavewriter Alpha 16 Implantable Pulse Generator Kit

Issue

Boston Scientific has received reports of undesired sensation and/or awareness of IPG turning off and on during IPG charging for patients with rechargeable SCS IPGs. Investigation of these reports determined that the device may reset (by design) due to detection of noise or interference during IPG charging. When the system resets, the patient's programmed stimulation turns off for approximately 10-15 seconds and then turns back on. When this occurs, some patients may perceive the sudden changes in stimulation. As soon as the reset is complete, the IPG resumes normal operation, including delivery of stimulation therapy. Boston Scientific has identified improvements (including an IPG firmware update for implanted devices) that will prevent this device reset behavior from occurring during IPG charging. Updating the IPG firmware eliminates the possibility of a coincident routine system check during IPG charging, thus preventing a potential system reset. This firmware update is available for any rechargeable SCS IPGs identified. Recall Start Date: July 17, 2024

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare
CategoryHealth products - Medical devices - Cardiovascular
CompaniesBoston Scientific Neuromodulation Corporation 25155 Rye Canyon Loop, Valencia, California, United States, 91355
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-75920
Background{}
CompaniesBoston Scientific Neuromodulation Corporation 25155 Rye Canyon Loop, Valencia, California, United States, 91355
AudiencesHealthcare
Source updated2024-08-14