Quick summary: VNS Therapy SenTiva Generators. Date: 2025-02-05.
VNS Therapy SenTiva Generators
2025-02-05
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: VNS Therapy SenTiva Generators. Date: 2025-02-05.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Complaints have been received where the SenTiva generator is reporting a loss of therapy or inability to make therapy changes with the VNS therapy programming system. A reed switch malfunction, resulting in the reed switch sticking in a closed position has been identified as the cause of these device events. Recall start date: January 10, 2025
What you should do
- Customers to be notified. Instructions to follow on response.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.