Public recall detail

Vivo 45 LS

2026-05-01

medium
Anaesthesiology Performance
Quick summary

Quick summary: Vivo 45 LS. Reason: Performance. Category: Anaesthesiology. Date: 2026-05-01.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: Vivo 45 LS. Reason: Performance. Category: Anaesthesiology. Date: 2026-05-01.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandBreas Medical Ab
ProductVivo 45 LS
IssueMedical devices - Performance
Recall classType II
IdentificationRA-82020

Affected products

ProductVivo 45 LS
LotS/N: 1R070KF to 1R100F9
Model230000
Serial numberS/N: 1R070KF to 1R100F9
DescriptionVivo 45 LS

Issue

In devices with the affected firmware versions, treatment in volume-controlled ventilation modes (VCV, VCV-SIMV and VCV-MPV) can, under specific trigger conditions, result in delivered tidal volume being reduced significantly below the set value.

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare
CategoryHealth products - Medical devices - Anaesthesiology
CompaniesBreas Medical AB Foretagsvagen 1, Molnlycke, Sweden, 43533
Recall date2026-04-17
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-82020
Background{}
CompaniesBreas Medical AB Foretagsvagen 1, Molnlycke, Sweden, 43533
AudiencesHealthcare
Source updated2026-05-01