Quick summary: Vitros® Systems. Date: 2024-03-07.
Vitros® Systems
2024-03-07
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Vitros® Systems. Date: 2024-03-07.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Some serial numbers of the Vitros® systems as listed above may experience an anomaly involving the sampling center, which may have been manufactured using a sub-optimal hardware part. Quidelortho confirmed a complaint where the sampling center mechanism unintentionally disengaged from the universal sample tray (tray), resulting in the misalignment of the intended sample container to be aspirated. From our investigation of this single occurrence, Quidelortho determined that if this disengagement and misalignment occurs, it may cause one or multiple patient samples to be aspirated from an unintended sample container within the same tray and report the results using the incorrect patient samples. Based on the investigation, Quidelortho estimates the likelihood of this occurrence to be approximately 0.00000035% (or about 1 in 2.88 million tests.) Recall start date: January 26, 2024
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.