Vercise Genus ™ R16 and R32 Implantable Pulse Generator (IPG) Kits

2024-08-12

medium
Recall Safety notice
Quick summary

Quick summary: Vercise Genus ™ R16 and R32 Implantable Pulse Generator (IPG) Kits. Date: 2024-08-12.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: Vercise Genus ™ R16 and R32 Implantable Pulse Generator (IPG) Kits. Date: 2024-08-12.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandBoston Scientific Neuromodulation Corporation
ProductVercise Genus ™ R16 and R32 Implantable Pulse Generator (IPG) Kits
IssueMedical devices - Performance
Recall classType I
IdentificationRA-75912

Affected products

ProductVercise Genus ™ R32 Implantable Pulse Generator (IPG) Kit
LotMore than 10 numbers, contact manufacturer.
ModelDB-1232
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionVercise Genus ™ R32 Implantable Pulse Generator (IPG) Kit
ProductVercise Genus ™ R16 Implantable Pulse Generator (IPG) Kit
LotMore than 10 numbers, contact manufacturer.
ModelDB-1216
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionVercise Genus ™ R16 Implantable Pulse Generator (IPG) Kit

Issue

Manufacturer is initiating an advisory to remind customers to follow steps outlined in the labeling/instructions for use to implant Vercise Deep Brain Stimulation (DBS) Implantable Pulse Generators (IPGs) within a subcutaneous pocket. Device header feedthrough (FT) wire break(s) have occurred in rechargeable Vercise Genus ™ DBS IPGs that were implanted submuscular in the pectoral location. Manufacturer has not received reports of FT wire breaks with Vercise Genus ™ DBS IPGs implanted in a subcutaneous pocket. Complete or partial breaks of device header FT wire(s) will prevent successful delivery of stimulation therapy, thus requiring removal/replacement of the device. Clinical observations of high monopolar impedances, undesired sensation, sudden loss of therapy, return of pre-implant symptoms and/or bluetooth connectivity challenges may be potential signals associated with FT wire break(s). Recall start date: August 8, 2024

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare
CategoryHealth products - Medical devices - Neurology
CompaniesBoston Scientific Neuromodulation Corporation 25155 Rye Canyon Loop, Valencia, California, United States, 91355
Published byHealth Canada
Recall classType I
Alert / recall typeHealth product recall
Identification numberRA-75912
Background{}
CompaniesBoston Scientific Neuromodulation Corporation 25155 Rye Canyon Loop, Valencia, California, United States, 91355
AudiencesHealthcare
Source updated2024-08-12