Public recall detail

Veradius Unity and BV Pulsera Mobile C-Arm System-9in. and 12in. Versions

2026-09-09

medium
Health products Health product recall
Quick summary

Quick summary: Veradius Unity and BV Pulsera Mobile C-Arm System-9in. and 12in. Versions. Reason: Health product recall. Category: Health products. Date: 2026-09-09.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: Veradius Unity and BV Pulsera Mobile C-Arm System-9in. and 12in. Versions. Reason: Health product recall. Category: Health products. Date: 2026-09-09.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandPHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
ProductVeradius Unity and BV Pulsera Mobile C-Arm System-9in. and 12in. Versions
IssueMedical devices - Performance
Recall classType II
IdentificationRA-82610

Affected products

ProductVeradius Unity
LotNot applicable.
Model718132
Serial numberNot applicable.
DescriptionVeradius Unity
ProductBV Pulsera Mobile C-Arm System-9in. And 12in. Versions
LotNot applicable.
Model718095
Serial numberNot applicable.
DescriptionBV Pulsera Mobile C-Arm System-9in. And 12in. Versions

Issue

Air bubbles may develop in the monoblock oil when there is a prolonged period between manufacture and installation, for example due to storage, transportation, or other pre-installation holding time. During a clinical procedure, when the user activates X-ray imaging, if an air bubble becomes trapped in the high-voltage area of the monoblock, there could be an internal electrical discharge (flashover) resulting in X-rays not being generated and consequently no image being displayed on the examination monitor. When this situation occurs, the system displays an error message indicating that the X-ray generator is not available. In some cases, restarting the system (which takes approximately 5 minutes) may restore X-ray generation. Philips' investigation indicates that this issue may occur during the early period of system use following a prolonged period between manufacture and installation. Based on the investigation, potentially affected systems have been identified as those for which more than 300 days elapsed between manufacture and installation and fewer than 250 days have elapsed since installation with continued system use, the monoblock oil temperature increases, allowing storage-related air bubbles to gradually dissipate over time.

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare
CategoryHealth products - Medical devices - Radiology
CompaniesPhilips Medical Systems Nederland B.V. Veenpluis 6, Best, Netherlands, 5684 PC
Recall date2026-09-01
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-82610
Background{}
CompaniesPhilips Medical Systems Nederland B.V. Veenpluis 6, Best, Netherlands, 5684 PC
AudiencesHealthcare
Source updated2026-09-09