Quick summary: Vanta/Sequentia Lt Clinician Programmer Application.
Vanta/Sequentia Lt Clinician Programmer Application
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Vanta/Sequentia Lt Clinician Programmer Application.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Connection issue for Vanta pain pc implantable neurostimulators when transplanted in the same patient. "Too many devices" error will appear from Clinician Programmer Application (CPA) when there is more than 1 Vanta ins detected by the clinician telemetry module. Current Vanta labeling does not include information with respect to what the "too many devices found" informational message means or how to troubleshoot it. Recall start date: August 4, 2022
What you should do
- Contact the manufacturer if you require additional information.