Quick summary: UniCel DxI 600 Access Immunoassay System-Total T4 Assay. Reason: Performance. Category: Chemistry. Date: 2026-03-19.
UniCel DxI 600 Access Immunoassay System-Total T4 Assay
2026-03-19
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: UniCel DxI 600 Access Immunoassay System-Total T4 Assay. Reason: Performance. Category: Chemistry. Date: 2026-03-19.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Access Total T4 calibrator lots (PN 33805) exhibit a negative bias on DxI 600/800 instruments, with slopes outside product specifications. In the worst-case scenario, there is a medium safety risk associated with this issue due to the potential to cause false low patient results.
What you should do
- Contact the manufacturer if you require additional information.