Affected products
Issue
Beckman identified an issue on UniCel DxI 600 & 800 Systems with linear slide pnp gantries running software versions 5.7 and up. During installs or upgrades, the Beckman identified an issue on UniCel DxI 600 & 800 Systems with linear slide pnp gantries running software versions 5.7 and up. During installs or upgrades, the pnp xmotor current incorrectly resets to 1.5a (instead of the required 2.0a) because the software overwrites the factory setting, causing increased pnp motion errors. current incorrectly resets to 1.5a (instead of the required 2.0a) because the software overwrites the factory setting, causing increased pnp motion errors. Recall start date: June 18, 2025