Quick summary: Trilogy EVO and EV300. Reason: Device compatibility. Category: Anaesthesiology. Date: 2026-02-13.
Trilogy EVO and EV300
2026-02-13
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Trilogy EVO and EV300. Reason: Device compatibility. Category: Anaesthesiology. Date: 2026-02-13.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Philips Respironics prohibits the use of non-pneumatic (e.g., vibrating mesh) nebulizers (FSN 2025-CC-SRC-020) with Trilogy EVO platform ventilators (including Trilogy EVO, Trilogy EVO O2, and Trilogy EV300 ventilators). The firm notifies clients about revisions to a prior field safety notice concerning nebulized aerosol deposition on the internal flow sensor (FSN 2024-CC-SRC-013 - flow sensor nebulized aerosol deposition)Philips Respironics is no longer permitting the use of non-pneumatic nebulizers with Trilogy EVO platform ventilators.These updates are necessary for ongoing safe and compliant functioning.Philips Respironics has issued a follow-up letter to Field Safety Notice (FSN) 2026‑CC‑SRC‑002, to provide updated guidance regarding nebulizer use with the Trilogy Evo Platform ventilators (Trilogy Evo, Trilogy Evo O₂, Trilogy EV300).*Updated Guidance:Non‑pneumatic nebulizers may continue to be used, provided that:they are positioned according to the updated nebulizer placement instructions, andthe warnings in the revised FSN are fully understood and followed.This update clarifies that non‑pneumatic nebulizers (e.g., vibrating mesh) may continue to be used with Trilogy Evo ventilators, if they are positioned correctly according to the updated placement instructions. This revised guidance supports continuity of care in situations where other nebulization options are limited. Philips Respironics will update the User Manual to include the new guidance on using non‑pneumatic nebulizers. A software update will also be released so the device interface clearly reflects that the guidance applies to both pneumatic and non‑pneumatic nebulizers. These updates will be included in a future release to limit repeated changes for users.NOTE: This update does not change the previously communicated information regarding:• Nebulized aerosol deposition on the internal flow sensor (FSN 2024-CC-SRC-013 – Flow Sensor Nebulized Aerosol Deposition)For further support, contact your local Philips representative.
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.
Starting...
Explore more
Related recalls
Similar recalls and safety notices.