Public recall detail

Tracoe Experc Percut Dilation Tracheostomy Set W/Trachatraumatic Inserter

2026-10-14

medium
Health products Health product recall
Quick summary

Quick summary: Tracoe Experc Percut Dilation Tracheostomy Set W/Trachatraumatic Inserter. Reason: Health product recall. Category: Health products. Date: 2026-10-14.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: Tracoe Experc Percut Dilation Tracheostomy Set W/Trachatraumatic Inserter. Reason: Health product recall. Category: Health products. Date: 2026-10-14.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandTracoe Medical GmbH
ProductTracoe Experc Percut Dilation Tracheostomy Set W/Trachatraumatic Inserter
IssueMedical devices - Performance
Recall classType II
IdentificationRA-82782

Affected products

ProductTracoe Experc Percut Dilation Tracheostomy Set W/Trachatraumatic Inserter
Lot1100026287
Model461-08-P
Serial number1100026287
DescriptionTracoe Experc Percut Dilation Tracheostomy Set W/Trachatraumatic Inserter
ProductTracoe Experc Percut Dilation Tracheostomy Set W/Trachatraumatic Inserter
Lot1100024157
Model424-08
Serial number1100024157
DescriptionTracoe Experc Percut Dilation Tracheostomy Set W/Trachatraumatic Inserter
ProductTracoe Experc Percut Dilation Tracheostomy Set W/Trachatraumatic Inserter
Lot1100024241
Model424-08
Serial number1100024241
DescriptionTracoe Experc Percut Dilation Tracheostomy Set W/Trachatraumatic Inserter
ProductTracoe Experc Percut Dilation Tracheostomy Set W/Trachatraumatic Inserter
Lot1100030719
Model424-08
Serial number1100030719
DescriptionTracoe Experc Percut Dilation Tracheostomy Set W/Trachatraumatic Inserter
ProductTracoe Experc Percut Dilation Tracheostomy Set W/Trachatraumatic Inserter
Lot1100022836
Model423-08
Serial number1100022836
DescriptionTracoe Experc Percut Dilation Tracheostomy Set W/Trachatraumatic Inserter
ProductTracoe Experc Percut Dilation Tracheostomy Set W/Trachatraumatic Inserter
Lot1100030666
Model423-08
Serial number1100030666
DescriptionTracoe Experc Percut Dilation Tracheostomy Set W/Trachatraumatic Inserter

Issue

Tracoe is issuing a Field Safety Notice as the application bears a very low risk that the silicone sleeve may detach from the guiding catheter when removing the insertion system of the tracheostomy tube. The sleeve may remain in the cannula and block it, requiring an immediate cannula change, or it may slip into the trachea, necessitating immediate bronchoscopy removal. The incidence rate is very low.

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare
CategoryHealth products - Medical devices - Anaesthesiology
CompaniesTracoe Medical GmbH Reichelsheimer Str.1/3, Nieder-Olm, Germany, 55268
Recall date2026-09-24
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-82782
Background{}
CompaniesTracoe Medical GmbH Reichelsheimer Str.1/3, Nieder-Olm, Germany, 55268
AudiencesHealthcare
Source updated2026-10-14