Quick summary: Tracheal Tube Reusable Introducer And Guides. Date: 2024-11-01.
Tracheal Tube Reusable Introducer And Guides
2024-11-01
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Tracheal Tube Reusable Introducer And Guides. Date: 2024-11-01.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Smiths Medical has identified a potential for ingress of fluid into the device during reprocessing. The ingress takes place at the rear of the device between the end cap and the tube. This could lead to staining of the device or allowing the fluid to remain in the device. In addition, the hypochlorite solution (200ppm) and the 4% acetic acid disinfectants recommended in the IFU may be inadequate according to the disinfection standards for this type of device. Recall Start Date: October 15, 2024
What you should do
- Contact the manufacturer if you require additional information.