Quick summary: Terumo Cardiovascular Procedure Kits. Date: 2024-03-28.
Terumo Cardiovascular Procedure Kits
2024-03-28
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Terumo Cardiovascular Procedure Kits. Date: 2024-03-28.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
CHS initiated this recall after receiving a notification from Terumo on their cardiovascular procedure kits that include high crack check valve. Terumo cardiovascular systems (Terumo CVS) has received six complaints whereby the pressure relief valve included in certain lots of cardiovascular procedure kits was reportedly opening at approximately 200-300mmhg versus 400mmhg. Recall start date: February 26, 2024
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.