Quick summary: Syncvision System.
Syncvision System
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Syncvision System.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
If prior to the IFR/FFR pullback on the intrasight system, FFR measurement(s) were performed within the same syncvision procedural session, the co-registered distal value will be different than the distal value displayed in the philips IFR/FFR system. This error will appear "insufficient data, distal segment is not co-registered" the risk of suboptimal patient treatment due to falsely elevated data on the initial trendline for the IFR/FFR co-registration would be greatest from inexperienced interventionalists. Recall start date: Dec 10, 2021
What you should do
- Contact the manufacturer if you require additional information.