Stryker Instruments System 8 and 9 Saws

2025-03-26

medium
Recall Safety notice
Quick summary

Quick summary: Stryker Instruments System 8 and 9 Saws. Date: 2025-03-26.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: Stryker Instruments System 8 and 9 Saws. Date: 2025-03-26.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandStryker Instruments, A Division Of Stryker Corporation
ProductStryker Instruments System 8 and 9 Saws
IssueMedical devices - Performance
Recall classType II
IdentificationRA-77182

Affected products

ProductSystem 8 Reciprocating Saw
Lot2409624823 1802305513
Model8206-000-000
Serial number2409624823 1802305513
DescriptionSystem 8 Reciprocating Saw
ProductSystem 9 Sternum Saw
LotNot applicable.
Model9207-000-000
Serial numberNot applicable.
DescriptionSystem 9 Sternum Saw
ProductSystem 9 Reciprocating Saw
LotMore than 10 numbers, contact manufacturer.
Model9206-000-000
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionSystem 9 Reciprocating Saw

Issue

There is a potential for the compression spring (0038-419-000) inside the saw's drivetrain assembly to disengage during blade loading and unloading. A disengaged spring disables the blade collar from springing back to its original home position, potentially leading to blade retention failure when loading a blade. Recall Start Date: March 14, 2025

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare
CategoryHealth products - Medical devices - Dental Health products - Medical devices - Orthopaedics Health products - Medical devices - General and plastic surgery
CompaniesStryker Instruments, A Division Of Stryker Corporation 1941 Stryker Way, Portage, Michigan, United States, 49002
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-77182
Background{}
CompaniesStryker Instruments, A Division Of Stryker Corporation 1941 Stryker Way, Portage, Michigan, United States, 49002
AudiencesHealthcare
Source updated2025-04-24

Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.