StealthStation™ S8 software application
2024-09-04
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: StealthStation™ S8 software application. Date: 2024-09-04.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Medtronic has identified the following two software anomalies in StealthStation™ S8 software application versions 1.3.0 and 1.3.2 that can occur within a spine procedure. 1. When lateral orientation radiographic images are taken using a navigation-enabled Ziehm Vision FD Vario 3D C-arm, the navigation orientation information appears flipped when displayed by the StealthStation™ S8 system. 2. When using CT + fluoroscopy images with a slice spacing thickness greater than 2 mm, the system does not display a pop up message indicating the slice spacing is not optimal. Navigation remains accurate in this situation. Recall start date: August 26, 2024
What you should do
- Contact the manufacturer if you require additional information.