Quick summary: STA – Liquid Anti-Xa. Date: 2025-07-25.
STA – Liquid Anti-Xa
2025-07-25
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: STA – Liquid Anti-Xa. Date: 2025-07-25.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Following internal tests, contamination of STA®-LIQUID ANTI-Xa by Stago Fibrinogen reagents was identified in UFH and LMWH applications on STA SATELLITE® or STA SATELLITE MAX® analyzers. Contamination occurs when one or more tests with the STA®-Fibrinogen 5 reagent are performed before the STA®-LIQUID ANTI-Xa test for heparin monitoring on an STA SATELLITE® or STA SATELLITE MAX® analyzer. This contamination causes an underestimation of heparin levels which could lead to a heparin treatment change and a subsequent overdose risk for the patient. Recall start date: July 21, 2025
What you should do
- Contact the manufacturer if you require additional information.
Additional official information