Quick summary: Sphere-9™ Catheter. Reason: Health product recall. Category: Health products. Date: 2026-07-27.
Sphere-9™ Catheter
2026-07-27
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Sphere-9™ Catheter. Reason: Health product recall. Category: Health products. Date: 2026-07-27.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
There have been forty-six (46) reported incident reports (none of which were from Canada) where patients with a Biotronik ICD experienced induced ventricular arrhythmia (VT and/or VF) following RF ablation of cardiac tissue using the Affera Sphere-9 catheter.Sphere-9 presents a higher current flow to the ICD or CRT-D which interacts with a safety feature built into Biotronik ICDs and CRT-Ds. The higher currents of the Sphere-9 catheter during RF application proceed to an unfavorable modulation of the coupled currents (approx. 40Hz frequency) that may be conducted via the implanted leads, and result in possible VT or VF induction.
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.