Quick summary: SOLTIVE™ Premium SuperPulsed Laser System. Date: 2024-09-25.
SOLTIVE™ Premium SuperPulsed Laser System
2024-09-25
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: SOLTIVE™ Premium SuperPulsed Laser System. Date: 2024-09-25.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Olympus received a complaint regarding the SOLTIVE™ Laser System preset laser settings. The term "bladder stone" was incorrectly translated in both Spanish and Portuguese to "kidney stone" (cálculo renal) on the systems' graphical user interface (GUI). As a result of the incorrect Spanish and Portuguese translation from "bladder stone" to "kidney stone" on the GUI, there is a potential to deliver an incorrect or unintended amount of energy to the patient's anatomy. This issue is specific to the preset laser settings and does not affect users who are not utilizing the preset option. Recall start date: September 12, 2024
What you should do
- Contact the manufacturer if you require additional information.
Additional official information