Public recall detail

Soft-Vu and Accu-Vu 4 FR Angiographic Catheters

2026-03-18

medium
Cardiovascular - Radiology Device compatibility - Performance
Quick summary

Quick summary: Soft-Vu and Accu-Vu 4 FR Angiographic Catheters. Reason: Device compatibility - Performance. Category: Cardiovascular - Radiology. Date: 2026-03-18.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: Soft-Vu and Accu-Vu 4 FR Angiographic Catheters. Reason: Device compatibility - Performance. Category: Cardiovascular - Radiology. Date: 2026-03-18.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandAngioDynamics, Inc.
ProductSoft-Vu and Accu-Vu 4 FR Angiographic Catheters
IssueMedical devices - Device compatibility Medical devices - Performance
Recall classType II
IdentificationRA-81769

Affected products

ProductSoft-Vu Angiographic Catheters: Berentstein: (Braided & Non)
LotMore than 10 numbers, contact manufacturer.
Model10714009
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionSoft-Vu Angiographic Catheters: Berentstein: (Braided & Non)
ProductSoft-Vu Angiographic Catheters: Berentstein: (Braided & Non)
LotMore than 10 numbers, contact manufacturer.
Model10714016
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionSoft-Vu Angiographic Catheters: Berentstein: (Braided & Non)
ProductAccu-Vu Sizing Catheter - Pigtail Flush
LotMore than 10 numbers, contact manufacturer.
Model13709804
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionAccu-Vu Sizing Catheter - Pigtail Flush
ProductSoft-Vu Angigraphic Catheters: Shepherd Hook (Braided & Non)
LotMore than 10 numbers, contact manufacturer.
ModelH787107100315
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionSoft-Vu Angigraphic Catheters: Shepherd Hook (Braided & Non)
ProductSoft-Vu Angiographic Catheters - Omni Flush
LotMore than 10 numbers, contact manufacturer.
Model10732301
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionSoft-Vu Angiographic Catheters - Omni Flush
ProductSoft-Vu Angiographic Catheters - Omni Flush
LotMore than 10 numbers, contact manufacturer.
Model10714001
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionSoft-Vu Angiographic Catheters - Omni Flush
ProductSoft-Vu Angiographic Flush Catheters: Pigtail
LotMore than 10 numbers, contact manufacturer.
Model10707103
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionSoft-Vu Angiographic Flush Catheters: Pigtail
ProductSoft-Vu Angiographic Flush Catheters: Pigtail
LotMore than 10 numbers, contact manufacturer.
Model10714002
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionSoft-Vu Angiographic Flush Catheters: Pigtail
ProductSoft-Vu Angiographic Selective Catheters: Cobra (Braided & Non)
LotMore than 10 numbers, contact manufacturer.
Model10714018
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionSoft-Vu Angiographic Selective Catheters: Cobra (Braided & Non)
ProductSoft-Vu Angiographic Selective Catheters: Cobra (Braided & Non)
LotMore than 10 numbers, contact manufacturer.
Model10714017
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionSoft-Vu Angiographic Selective Catheters: Cobra (Braided & Non)
ProductSoft-Vu Angiographic Catheters - Kumpe
LotMore than 10 numbers, contact manufacturer.
Model10714021
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionSoft-Vu Angiographic Catheters - Kumpe
ProductSoft-Vu Angiographic Catheters - Kumpe
LotMore than 10 numbers, contact manufacturer.
Model10714014
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionSoft-Vu Angiographic Catheters - Kumpe
ProductSoft-Vu Angiographic Catheters - Kumpe
LotMore than 10 numbers, contact manufacturer.
Model10734301
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionSoft-Vu Angiographic Catheters - Kumpe

Issue

AngioDynamics, Inc., is conducting a voluntary medical device field safety corrective action or recall to the end-user level based on the non-conformance of a molded hub in specific models and lots of Soft-Vu and Accu-Vu 4 fr angiographic catheters. The affected Soft-Vu and Accu-Vu 4 fr angiographic catheters being recalled (removed) contain a manufacturing defect which may prevent the appropriately sized guidewire from passing through the inner diameter (ID) of the catheter hub. AngioDynamics has become aware of the non-conformance by way of multiple complaints associated with the affected angiographic catheter devices; however, there have been no complaints or events that occurred in Canada. There are no reports of patient injury as a result of this issue.

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare Industry
CategoryHealth products - Medical devices - Cardiovascular Health products - Medical devices - Radiology
CompaniesAngioDynamics, Inc. 603 Queensbury Avenue, Queensbury, New York, United States, 12804
Recall date2026-03-03
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-81769
Background{}
CompaniesAngioDynamics, Inc. 603 Queensbury Avenue, Queensbury, New York, United States, 12804
AudiencesHealthcare Industry
Source updated2026-03-18

Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.

Starting...

Related recalls

Similar recalls and safety notices.

Loading related recalls...