Sight Olo-E1 & Sight Olo-U1

medium
Recall Safety notice
Quick summary

Quick summary: Sight Olo-E1 & Sight Olo-U1.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: Sight Olo-E1 & Sight Olo-U1.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandS.D. Sight Diagnostics Ltd.
ProductSight Olo-E1 & Sight Olo-U1
IssueMedical devices - Performance
Recall classType II
IdentificationRA-64349

Affected products

ProductSight Olo-E1
Lotversion 2.63.1.1 version 2.63.1 version 2.63
ModelOLO-E1
Serial numberversion 2.63.1.1 version 2.63.1 version 2.63
DescriptionSight Olo-E1
ProductSight Olo-U1
Lotversion 2.63.1.1 version 2.63.1 version 2.63
ModelOLO-U1
Serial numberversion 2.63.1.1 version 2.63.1 version 2.63
DescriptionSight Olo-U1

Issue

The reference ranges of MCH, MCHC and RDW are mixed in the child (2-<12 years old) demographics. if the user has not overridden them, they will be incorrectly displayed. Recall start date: June 14, 2022

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare Industry
CategoryHealth products - Medical devices - Haematology
CompaniesS.D. Sight Diagnostics Ltd. 23 Menachem Begin Rd., Tel Aviv, Israel, 6618356
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-64349
Background{}
CompaniesS.D. Sight Diagnostics Ltd. 23 Menachem Begin Rd., Tel Aviv, Israel, 6618356
AudiencesHealthcare Industry
Source updated2022-07-11

Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.