Quick summary: Siemens Biograph mMR System.
Siemens Biograph mMR System
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Siemens Biograph mMR System.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Siemens Healthcare GmbH became aware of this issue as part of a customer complaint investigation. It has been identified that this issue can occur when certain parameters (Echo spacings 0.50 MS; 0.51 MS, and 0.52 MS) are used in EPI (Echo-Planar Imaging), sequences will excite high internal stress inside the gradient coil assembly. This may lead to cracks and damage to the gradient coil. Recall start date: March 15, 2024
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.