Public recall detail

Sevoflurane, Desflurane, and Isoflurane Electronic Vaporizer Material Package

2024-08-20

medium
Recall Safety notice
Quick summary

Quick summary: Sevoflurane, Desflurane, and Isoflurane Electronic Vaporizer Material Package. Date: 2024-08-20.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: Sevoflurane, Desflurane, and Isoflurane Electronic Vaporizer Material Package. Date: 2024-08-20.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandShenzhen Mindray Bio-Medical Electronics Co., Ltd.
ProductSevoflurane, Desflurane, and Isoflurane Electronic Vaporizer Material Package
IssueMedical devices - Performance
Recall classType II
IdentificationRA-75948

Affected products

ProductSevoflurane (Quik-Fil) Electronic Vaporizer Material Package
LotMore than 10 numbers, contact manufacturer.
Model115-066755-00
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionSevoflurane (Quik-Fil) Electronic Vaporizer Material Package
ProductDesflurane (Saf-T-Fill) Electronic Vaporizer Material Package
LotMore than 10 numbers, contact manufacturer.
Model115-066759-00
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionDesflurane (Saf-T-Fill) Electronic Vaporizer Material Package
ProductIsoflurane (Safety Filling Adapter) Electronic Vaporizer Material Package
LotMore than 10 numbers, contact manufacturer.
Model115-066758-00
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionIsoflurane (Safety Filling Adapter) Electronic Vaporizer Material Package

Issue

The anesthesia system is a device used to administer a general inhalation anesthetic continuously or intermittently to a patient and maintain their ventilation. The A9 anesthesia system utilizes electronic vaporizers (V90) for anesthetic agent delivery. Mindray has become aware of a potential issue with select batches of V90 electronic vaporizers. The V90 electronic vaporizer anesthetic agent output may be higher than the set amount due to a component within the vaporizer causing an incomplete seal in some cases. Should this issue occur, the A9 anesthesia system will generate alarms associated with abnormal vaporizer output. We recommend that the user use the part number reference to identify the license and the device identifier. Recall start date: July 22, 2024

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare
CategoryHealth products - Medical devices - Anaesthesiology
CompaniesShenzhen Mindray Bio-Medical Electronics Co., Ltd. Mindray Building, Suite 3b, Shenzhen, Guangdong, China, 518057
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-75948
Background{}
CompaniesShenzhen Mindray Bio-Medical Electronics Co., Ltd. Mindray Building, Suite 3b, Shenzhen, Guangdong, China, 518057
AudiencesHealthcare
Source updated2024-11-15

Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.

Starting...

Related recalls

Similar recalls and safety notices.

Loading related recalls...