Quick summary: Roadrunner Guidewire.
Roadrunner Guidewire
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Roadrunner Guidewire.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Cook Medical identified that samples from the affected device lots did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration testing. This failure was traced to material that is supplied to Cook Medical from an external supplier. Recall start date: July 19, 2022
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.