Quick summary: Revaclear 400 Dialyzer; Revaclear 300 Dialyzer. Date: 2022-12-02.
Revaclear 400 Dialyzer; Revaclear 300 Dialyzer
2022-12-02
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Revaclear 400 Dialyzer; Revaclear 300 Dialyzer. Date: 2022-12-02.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Baxter Corporation is communicating two separate, but associated, issues related to Revaclear dialyzers. The first issue is linked to reports of internal blood leaks being caused by a twisted gasket in the dialyzer, prompting Baxter to recall the impacted lots. The second issue is linked to reports of polyurethane displacement in the dialyzer, prompting Baxter to issue a safety alert for the impacted lots. Recall start date: November 24, 2022
What you should do
- Contact the manufacturer if you require additional information.