Quick summary: Respironics Ventilator.
Respironics Ventilator
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Respironics Ventilator.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
All V60/V60 plus and V680 units have been identified to have potential issue that could affect the main electrical circuit ("35V rail") powering the ventilator and alarm. In some cases, this issue may result in either one of the following scenarios: 1. The ventilator ceases to operate, activating both visual and audible alarms. 2. The ventilator ceases to operate and does not activate either a visual or audible alarm causing a "silent shutdown". Recall start date: Mar 17, 2022
What you should do
- Contact the manufacturer if you require additional information.