Quadrox-I Adult Oxygenator W/Integrated Hardshell Reservoir
2024-08-30
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Quadrox-I Adult Oxygenator W/Integrated Hardshell Reservoir. Date: 2024-08-30.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Maquet Cardiopulmonary's monthly inspection of oxygenators for the presence of the Softline coating revealed that two out of ten oxygenators tested did not pass. The Softline coating is a vital component in the biocompatibility of blood contact components used in extracorporeal circulation. This means that the oxygenators installed in the products listed above do not have the Softline coating. The issue is limited to one lot number in Canada. Recall Start Date: August 20, 2024
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.