Public recall detail

QIAcube Connect MDX

medium
Recall Safety notice
Quick summary

Quick summary: QIAcube Connect MDX.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: QIAcube Connect MDX.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandQIAGEN GMBH
ProductQIAcube Connect MDX
IssueMedical devices - Software defect
Recall classType II
IdentificationRA-63672

Affected products

ProductQIAcube Connect MDX
LotS/N: 30011 S/N: 30036
Model9003070
Serial numberS/N: 30011 S/N: 30036
DescriptionQIAcube Connect MDX

Issue

The information detailed on the screen when the info icon is clicked is incorrect and should state that the volume is exact, not a minimal volume. This could cause a user to pipette an incorrect volume which could lead to mistaken results or sample loss. The main screen and user manual are correct but this supplemental information screen could confuse a user. Discovered internally at Qiagen. Recall start date: 2021-11-18

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceGeneral public Health professionals Hospitals
CategoryHealth products - Medical devices
CompaniesQiagen Gmbh Qiagen Strasse 1, Hilden Germany, 40724
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-63672
Background{}
CompaniesQiagen Gmbh Qiagen Strasse 1, Hilden Germany, 40724
AudiencesGeneral public Health professionals Hospitals
Source updated2021-12-02

Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.