Quick summary: QAngio XA 3D. Date: 2024-12-24.
QAngio XA 3D
2024-12-24
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: QAngio XA 3D. Date: 2024-12-24.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
For certain Siemens XA images, the isocenter calibration value is calculated incorrectly due to an incorrect value for distance source to isocenter that is saved by Siemens in the standard dicom attribute (0018,1111). The correct value is stored in a Siemens private dicom attribute (0021, xx17, "Siemens sms-ax acq 1.0") in the dicom data. Incorrect calibration can lead to errors in the measurements provided by qAngio XA 3D. Recall start date: December 16, 2024
What you should do
- Contact the manufacturer if you require additional information.