Quick summary: Pushtracker E2 & E3 Utilized With Smartdrive MX2+ Power Assist Device. Date: 2023-05-19.
Pushtracker E2 & E3 Utilized With Smartdrive MX2+ Power Assist Device
2023-05-19
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Pushtracker E2 & E3 Utilized With Smartdrive MX2+ Power Assist Device. Date: 2023-05-19.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Through design and development work, Max Mobility has identified a software issue with the application. When multiple processes are running on the watch's central processing unit, the application may crash unexpectedly. If this happens, the motor on the power assist device continues to run and the user may not be able to stop the device using tap gestures. Recall start date: May 2, 2023
What you should do
- Contact the manufacturer if you require additional information.