Provido, Proveo 8 and M530 OHX Microscopes

2023-09-28

medium
Recall Safety notice
Quick summary

Quick summary: Provido, Proveo 8 and M530 OHX Microscopes. Date: 2023-09-28.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: Provido, Proveo 8 and M530 OHX Microscopes. Date: 2023-09-28.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandLeica Microsystems Schweiz AG
ProductProvido, Proveo 8 and M530 OHX Microscopes
IssueMedical devices - Performance
Recall classType II
IdentificationRA-74338

Affected products

ProductM530 OHX – Optics Carrier
Lot10448737
Model291122001 11221001 281122001 140223001 60422002 140222002 301122002 230822001
Serial number10448737
DescriptionM530 OHX – Optics Carrier
ProductLeica Proveo 8 Ophthalmic Microscope System
Lot10449058
ModelMore than 10 numbers, contact manufacturer.
Serial number10449058
DescriptionLeica Proveo 8 Ophthalmic Microscope System
ProductProvido
Lot10448976 10448950
Model110722001 130722001 140722001
Serial number10448976 10448950
DescriptionProvido

Issue

Complaint trending revealed an adverse trend of complaints stating that a LMS surgical operating microscope unexpectedly shut down caused by a defective 24VDC power supply which is used inside the surgical operating microscopes Provido, Proveo 8 and M530 OHX. Based on the results of the investigations there is a reasonable probability that the MOSFET in the power supply unit of these microscopes will overheat and become defective. Recall Start Date: August 30, 2023

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare
CategoryHealth products - Medical devices Health products - Medical devices - Ear, nose and throat Health products - Medical devices - General and plastic surgery
CompaniesLeica Microsystems (Schweiz) Ag Max Schmidheiny-Strasse 201, Heerbrugg, St. Gallen, Switzerland, 9435
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-74338
Background{}
CompaniesLeica Microsystems (Schweiz) Ag Max Schmidheiny-Strasse 201, Heerbrugg, St. Gallen, Switzerland, 9435
AudiencesHealthcare
Source updated2023-09-28