Quick summary: Provido, Proveo 8 and M530 OHX Microscopes. Date: 2023-09-28.
Provido, Proveo 8 and M530 OHX Microscopes
2023-09-28
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Provido, Proveo 8 and M530 OHX Microscopes. Date: 2023-09-28.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Complaint trending revealed an adverse trend of complaints stating that a LMS surgical operating microscope unexpectedly shut down caused by a defective 24VDC power supply which is used inside the surgical operating microscopes Provido, Proveo 8 and M530 OHX. Based on the results of the investigations there is a reasonable probability that the MOSFET in the power supply unit of these microscopes will overheat and become defective. Recall Start Date: August 30, 2023
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.