Quick summary: pro+ Flat Deck Wide W/O X-Ray. Reason: Health product recall. Category: Health products. Date: 2026-08-27.
pro+ Flat Deck Wide W/O X-Ray
2026-08-27
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: pro+ Flat Deck Wide W/O X-Ray. Reason: Health product recall. Category: Health products. Date: 2026-08-27.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
The UDI label for model p7924a13 incorrectly states the product description as "Pro+ 40 MRS surface with x-ray" instead of the actual configuration, which is "Pro+ 36 MRS surface." Model p7924a13 should be used with 36" width bed frames and not 40" width bed frames. These discrepancies in labeling do not align with the approved product or product specifications and may lead to incorrect product identification or selection, potentially resulting in use with an incompatible bed frame.
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.