Quick summary: Powerheart AED G5 Semi-Automatic and Powerheart AED G5 Semi-Automatic With CPR Assist. Date: 2024-06-17.
Powerheart AED G5 Semi-Automatic and Powerheart AED G5 Semi-Automatic With CPR Assist
2024-06-17
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Powerheart AED G5 Semi-Automatic and Powerheart AED G5 Semi-Automatic With CPR Assist. Date: 2024-06-17.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
The G5 Semi-Automatic AED is shipped, with a protective film over its front panel, to protect the screen and shock button from cosmetic damage during shipping. It has come to manufacturer’s attention that customers may not be removing the protective film during deployment of the product. If the protective film is left adhered to the front bezel it may prevent the user from actuating the shock button. This may lead to a delay or prevent delivery of defibrillation therapy to a victim suffering from sudden cardiac arrest. Recall start date: June 10, 2024
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.