Public recall detail

Philips Respironics Trilogy EV300 and Evo ventilator

2023-04-26

medium
Recall Safety notice
Quick summary

Quick summary: Philips Respironics Trilogy EV300 and Evo ventilator. Date: 2023-04-26.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: Philips Respironics Trilogy EV300 and Evo ventilator. Date: 2023-04-26.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandRespironics Inc. Also Trading As Philips Rs North America Llc
ProductPhilips Respironics Trilogy EV300 and Evo ventilator
IssueMedical devices - Performance
Recall classType II
IdentificationRA-73573

Affected products

ProductTrilogy EV300
LotAll lots.
ModelCA2200X12B
Serial numberAll lots.
DescriptionTrilogy EV300
ProductTrilogy EVO
LotAll lots.
ModelCA2110X12B
Serial numberAll lots.
DescriptionTrilogy EVO

Issue

Philips Respironics has detected environmental contamination (e.g. external dust and dirt) in the air path of some devices that have been returned from the field. Extended exposure to these environmental contaminants can lead to buildup of particulate on a flow sensor within the device. This may cause the device to inaccurately deliver pressure or volume/flow. Recall start date: March 28, 2023

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare Industry
CategoryHealth products - Medical devices - Anaesthesiology
CompaniesRespironics Inc. Also Trading As Philips Rs North America LLC 1001 Murry Ridge Lane, Murrysville, Pennsylvania, United States, 15668
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-73573
Background{}
CompaniesRespironics Inc. Also Trading As Philips Rs North America LLC 1001 Murry Ridge Lane, Murrysville, Pennsylvania, United States, 15668
AudiencesHealthcare Industry
Source updated2023-04-26

Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.