Perfusor® Space® Syringe Pump

2024-06-17

medium
Recall Safety notice
Quick summary

Quick summary: Perfusor® Space® Syringe Pump. Date: 2024-06-17.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: Perfusor® Space® Syringe Pump. Date: 2024-06-17.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandB. Braun Melsungen AG
ProductPerfusor® Space® Syringe Pump
IssueMedical devices - Performance
Recall classType II
IdentificationRA-75684

Affected products

ProductSpace Infusion System - Perfusor® Space® Infusion Syringe Pump
LotMore than 10 numbers, contact manufacturer.
Model8713030U 8713031U
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionSpace Infusion System - Perfusor® Space® Infusion Syringe Pump
ProductSpace Infusion System - PCA-Kit
LotMore than 10 numbers, contact manufacturer.
Model8713080U
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionSpace Infusion System - PCA-Kit
ProductSpace Infusion System - Perfusor® Space® ACM Pump
LotMore than 10 numbers, contact manufacturer.
Model8713040U
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionSpace Infusion System - Perfusor® Space® ACM Pump
ProductPerfusor® Space®
LotMore than 10 numbers, contact manufacturer.
Model8713030C 8713032U
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionPerfusor® Space®

Issue

Cardinal Health began distributing Monoject™ syringes branded as "Cardinal Health Monoject™ syringes" however, the new syringes have different dimensions than prior Covidien Monoject™ syringes. The dimensional changes made to the Cardinal Health Monoject™ syringes, when used with syringe pumps, may result in recognition, compatibility, and pump performance issues, such as overdose, underdose, delay in therapy, delay in occlusion alarms, and delay in feeding. The Perfusor® space syringe pump IFU's identify Monoject™ syringes as an option for use with the pumps. Recall start date: June 4, 2024

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare
CategoryHealth products - Medical devices - General hospital and personal use
CompaniesB. Braun Melsungen AG Carl-Braun-Str. 1, Melsungen, Germany, 34212
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-75684
Background{}
CompaniesB. Braun Melsungen AG Carl-Braun-Str. 1, Melsungen, Germany, 34212
AudiencesHealthcare
Source updated2024-06-17