Quick summary: Paradigm and MiniMed 600 and 700 series pumps. Reason: Performance. Category: General hospital and personal use. Date: 2026-02-26.
Paradigm and MiniMed 600 and 700 series pumps
2026-02-26
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Paradigm and MiniMed 600 and 700 series pumps. Reason: Performance. Category: General hospital and personal use. Date: 2026-02-26.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
As part of the testing for the Alternate Controller Enabled (ACE) device submission, the R&D team tested the Minimed 780g Insulin Pump at elevated and lowered height conditions relative to the infusion site per the engineering test report. The results of the testing showed that there is an unintended over or under delivery of insulin that occurs when the 780g pump is either elevated or lowered relative to the infusion site.
What you should do
- Contact the manufacturer if you require additional information.