Quick summary: ORTHO VISION Analyzer. Date: 2024-11-22.
ORTHO VISION Analyzer
2024-11-22
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: ORTHO VISION Analyzer. Date: 2024-11-22.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
During an internal investigation Quidelortho™ identified a potential for a false positive test result due to carryover of ORTHO Sera Anti-D (DVI) on vision analyzers. Recall Start Date: November 8, 2024
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.
Starting...
Explore more
Related recalls
Similar recalls and safety notices.