Public recall detail

Origio Sperm Wash

2024-03-07

medium
Recall Safety notice
Quick summary

Quick summary: Origio Sperm Wash. Date: 2024-03-07.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: Origio Sperm Wash. Date: 2024-03-07.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandOrigio A/S
ProductOrigio Sperm Wash
IssueMedical devices - Performance
Recall classType II
IdentificationRA-75188

Affected products

ProductOrigio Sperm Wash
Lot230922-018009
Model84055060D
Serial number230922-018009
DescriptionOrigio Sperm Wash

Issue

During testing it was found that one of the batches, had an insufficient concentration of sodium pyruvate - only 0.002 mg/ml compared to a specified concentration of 0.046-0.137 mg/ml (1.5-4.3% of the specified amount). Part of this formulation batch had also been used to create three lots of commercially released sperm wash medium. Recall start date: January 22, 2024

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare
CategoryHealth products - Medical devices - Obstetrics and gynaecology
CompaniesOrigio A/S Knardrupvej 2, Malov, Denmark, 2760
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-75188
Background{}
CompaniesOrigio A/S Knardrupvej 2, Malov, Denmark, 2760
AudiencesHealthcare
Source updated2024-03-07

Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.