Quick summary: OptiSite and Fem-Flex II Arterial Perfusion Cannulas. Date: 2025-06-18.
OptiSite and Fem-Flex II Arterial Perfusion Cannulas
2025-06-18
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: OptiSite and Fem-Flex II Arterial Perfusion Cannulas. Date: 2025-06-18.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Edwards has confirmed a small number of occurrences impacting the OptiSite Arterial Perfusion Cannula in which a 3mm to 4mm section of wire, from the wire-reinforcement coil at the cannula tip, was found to be released from the cannula body. No customer complaints were reported in Canada. As certain femoral arterial cannula models are made of the same components as the impacted OptiSite Arterial Perfusion Cannula, the scope of this notice includes these femoral arterial cannula models. Recall start date: June 3, 2025
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.
Starting...
Explore more
Related recalls
Similar recalls and safety notices.