Quick summary: Optima Mr430s 1.5t - Main Unit.
Optima Mr430s 1.5t - Main Unit
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Optima Mr430s 1.5t - Main Unit.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
GE healthcare (GEHC) MR systems could potentially have a cryogen ventilation system that does not meet the venting requirements. Failure to have proper venting could present a safety issue if the cryogen gas is vented into the MR room during a magnet quench, potentially depriving the room of oxygen. Recall start date: Jan 4, 2022
What you should do
- Contact the manufacturer if you require additional information.