OmniaSecure™ MRI SureScan™

2026-09-23

medium
Health products Health product recall
Quick summary

Quick summary: OmniaSecure™ MRI SureScan™. Reason: Health product recall. Category: Health products. Date: 2026-09-23.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: OmniaSecure™ MRI SureScan™. Reason: Health product recall. Category: Health products. Date: 2026-09-23.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandMEDTRONIC INC.
ProductOmniaSecure™ MRI SureScan™
IssueMedical devices - Performance
Recall classType II
IdentificationRA-82665

Affected products

ProductOmniaSecure™ MRI SureScan™
LotMore than 10 numbers, contact manufacturer.
Model3930M79
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionOmniaSecure™ MRI SureScan™
ProductOmniaSecure™ MRI SureScan™
LotMore than 10 numbers, contact manufacturer.
Model3930M74
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionOmniaSecure™ MRI SureScan™

Issue

The model 3930M is a transvenous DF4 lead labelled for use in the RV and left bundle branch area (LBBA). It uses a fixed helix design, so the lead must be rotated as a whole for fixation. It's design transfers torque from the connector and lead body to the distal tip, supporting deep septal placement. To support this function, the lead incorporates interlocking mechanical features between the helix and tip to ring (tr) spacer, along with an adhesive bond applied during manufacturing. The adhesive is dispensed manually in a three-step process. All returned product analysis and some initial pho returns showed inadequate adhesive coverage at the tr spacer to helix interface. Current investigation indicates that operator technique and inspection contribute to the root cause. Insufficient adhesive at the interface between the driveshaft assembly and tr spacer causes a weak bond. Torque applied during the implant procedure may cause the adhesive bond to break. The helix then can separate from the TR spacer, referred to as driveshaft separation, while maintaining the overall lead connection with the cable. If separation occurs and the lead is implanted, the helix and cable assembly are unsupported, and lead fracture may occur over time.

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare
CategoryHealth products - Medical devices - Cardiovascular
CompaniesMedtronic Inc. 710 Medtronic Parkway N.E., Minneapolis, Minnesota, United States, 55432
Recall date2026-09-17
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-82665
Background{}
CompaniesMedtronic Inc. 710 Medtronic Parkway N.E., Minneapolis, Minnesota, United States, 55432
AudiencesHealthcare
Source updated2026-09-23

Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.

Starting...

Related recalls

Similar recalls and safety notices.

Loading related recalls...