Quick summary: Olympus Medical PK/PKS/Everest Cutting Forceps. Reason: Performance. Category: General and plastic surgery. Date: 2026-03-17.
Olympus Medical PK/PKS/Everest Cutting Forceps
2026-03-17
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Olympus Medical PK/PKS/Everest Cutting Forceps. Reason: Performance. Category: General and plastic surgery. Date: 2026-03-17.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Olympus identified inadequate supplier welding validation on several 5 mm cutting forceps models, which may cause jaw breakage during clinical use. Due to this defect, Olympus is requesting customers to return all affected products to prevent potential safety risks and ensure proper device performance.