Quick summary: Olympus HF Cable. Reason: Performance. Category: Gastroenterology and urology. Date: 2026-03-24.
Olympus HF Cable
2026-03-24
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Olympus HF Cable. Reason: Performance. Category: Gastroenterology and urology. Date: 2026-03-24.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Olympus received a complaint involving a thermal event with an HF cable during a procedure, causing a user burn. Investigation showed the device was used beyond IFU-defined service life, leading to insulation wear, corrosion, and failure. A four-year review identified 109 complaints, with no serious injuries and 17 minor burns.
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.