Quick summary: Olympus Air/Water Valves. Date: 2023-07-27.
Olympus Air/Water Valves
2023-07-27
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Olympus Air/Water Valves. Date: 2023-07-27.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Olympus discovered the MAJ-1444 may become damaged and result in loss of the one-way valve functionality by repeated automated endoscope reprocessing in Olympus OER machines. Specifically, when the MAJ-1444 is put into the washing case with other accessories of the ultrasonic endoscope, such as suction valve MAJ-1443, which may collide against each other causing damage. If the MAJ-1444 loses one-way valve functionality, body fluid could backflow into the air/water channel of the ultrasonic endoscopes during the procedure. There is no check that can be performed by users which evaluates one-way valve functionality. Therefore, a risk to patient health exists: 1) if the body fluid remains in the channel of the ultrasonic endoscopes following reprocessing, and 2) if the contaminated ultrasonic endoscope is used for the next patient without realizing the body fluid during the inspection, then a contaminated ultrasonic endoscope channel can cause transmission of infection. Recall Start Date: July 13, 2023
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.
Starting...
Explore more
Related recalls
Similar recalls and safety notices.